Academy / Field fault notes / Dräger Savina
Dräger · Savina

Dräger Savina: repeated flow sensor failures

What happened

Flow sensors kept failing one after another; the engineers reported about 30 sensors used instead of the yearly stock, while replacing the sensor cable changed nothing.

What was checked

Cause

Moisture, not electronics: with the humidifier and nebuliser in use and no filters after the valves and the nebuliser line, water droplets reached the flow sensor and destroyed it. The engineers judged this the probable cause in this particular case.

The cause is probable and was not fully confirmed.

What was done

No sensor repair involved — the fix was the setup: fit a bacterial filter (not a heat-and-moisture exchanger) on the expiratory side, restore all missing filters in the circuit, brief the staff on correct use with humidifier and nebuliser, and keep the unit under observation. Since the outcome was probable, monitoring sensor consumption afterwards was recommended.

Safety

Used flow sensors and breathing circuits are contaminated with patient secretions — handle and dispose of them as biohazard, and complete the self-test and electrical safety checks before returning the ventilator to a patient.

Lesson: When flow sensors fail in batches, check what the humidifier and nebuliser are sending into the circuit and whether the filters are actually in place before ordering more sensors.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.