Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Savina Select
Dräger · Savina Select

Dräger Savina Select: O2 sensor calibration needs a high flow the concentrators cannot deliver

What happened

The ventilator periodically requested an oxygen sensor calibration. The calibration procedure draws a high O2 flow, but the site runs only on oxygen concentrators that cannot cover that flow, and no cylinder was at hand. The unit was otherwise usable.

What was checked

Cause

The engineers attributed the repeated calibration requests to the unit being left off mains for long stretches: the battery drained and the heated oxygen sensor cooled down, which triggers a recalibration request. They treated this as the probable cause — it was not confirmed with measurements.

The cause is probable and was not fully confirmed.

What was done

A single calibration from a cylinder works, and the unit then runs on concentrators again. Keep in mind that the next auto-calibration will demand the same high flow, so a cylinder has to be arranged for that moment. Keeping the ventilator permanently on mains power, so the sensor heater stays warm, appeared to prevent the calibration request altogether — that was the probable outcome in this case. No parts were involved.

Safety

Calibration runs high oxygen flows, so keep oil, grease and ignition sources away, and complete the standard pre-use verification and the electrical safety test before the ventilator goes back to patients.

Lesson: A repeated calibration prompt on a heated oxygen sensor is often a power-habit issue rather than a sensor fault — keep the ventilator on mains and confirm the oxygen source needed for calibration before you start it.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes