Drager · Vista 120

Dräger Vista 120: endless reboot loop

What happened

The monitor keeps restarting endlessly and never settles into normal operation; no fault codes are shown.

What was checked

Cause

In this case the cause was never finally identified. The engineers treated it as probable that the loop came from an unstable supply rail, a sticking button or a faulty touchscreen holding the logic busy, or a poor connection in a ribbon cable — any of these can keep restarting the unit.

The cause is probable and was not fully confirmed.

What was done

Probable, not confirmed in this case: verify the power rails under load, re-seat and inspect the ribbon cables and connectors, then perform a reset to default settings — the reset is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Unplug the monitor from the mains and let the internal capacitors discharge before opening it, and run a full electrical safety check (protective earth continuity and leakage currents) before the monitor is returned to patient use.

Lesson: Treat a boot loop as a clue, not a verdict: rule out stuck buttons, peripherals and bad connections first, measure the supply rails at the moment of failure, and only then suspect the board.