Dräger Vista 120: endless reboot loop
What happened
The monitor keeps restarting endlessly and never settles into normal operation; no fault codes are shown.
What was checked
- Check that no front-panel button is sticking or held, and disconnect the touchscreen to see if the loop stops.
- Open the unit and watch the reported main rails — 12 V, 5 V and 3.3 V — at the exact moment of the restart, looking for a dip or dropout.
- Inspect flat cables and plug-in connectors for poor contact, oxidation or a half-seated plug.
Cause
In this case the cause was never finally identified. The engineers treated it as probable that the loop came from an unstable supply rail, a sticking button or a faulty touchscreen holding the logic busy, or a poor connection in a ribbon cable — any of these can keep restarting the unit.
The cause is probable and was not fully confirmed.
What was done
Probable, not confirmed in this case: verify the power rails under load, re-seat and inspect the ribbon cables and connectors, then perform a reset to default settings — the reset is a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Unplug the monitor from the mains and let the internal capacitors discharge before opening it, and run a full electrical safety check (protective earth continuity and leakage currents) before the monitor is returned to patient use.
Lesson: Treat a boot loop as a clue, not a verdict: rule out stuck buttons, peripherals and bad connections first, measure the supply rails at the moment of failure, and only then suspect the board.