Dräger · Zeus

Dräger Zeus: dust in the air/O2 flow sensors

What happened

The machine kept needing recalibration, and dust contamination of the air and O2 flow sensors was suspected.

What was checked

Cause

Contamination, in this one case only: the engineers found a layer of dust on the inner wall of the air flow sensor, and the air inlet filter was clogged and had turned grey. This is not a general rule.

The cause is probable and was not fully confirmed.

What was done

The engineers blew the flow sensors clean, cleaned the inlet filter and refitted it. Recalibration was recommended afterwards — a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable, so confirm the readings stay stable after calibration.

Safety

Decontaminate the machine and disconnect it from mains power and gas supplies before removing panels, and before return to use run the full leak test and self-test to confirm the cleaned flow sensors read correctly.

Lesson: A clogged inlet filter eventually lets dust through to the flow sensors, so when an anesthesia machine keeps drifting out of calibration, check and service the filter first.