Dräger Zeus: dust in the air/O2 flow sensors
What happened
The machine kept needing recalibration, and dust contamination of the air and O2 flow sensors was suspected.
What was checked
- Rear cover removed to reach the flow sensor area
- Air and O2 flow sensors inspected for dust build-up
- O2 flow sensor opened up for a closer inspection
- Air inlet filter pulled and checked
Cause
Contamination, in this one case only: the engineers found a layer of dust on the inner wall of the air flow sensor, and the air inlet filter was clogged and had turned grey. This is not a general rule.
The cause is probable and was not fully confirmed.
What was done
The engineers blew the flow sensors clean, cleaned the inlet filter and refitted it. Recalibration was recommended afterwards — a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable, so confirm the readings stay stable after calibration.
Safety
Decontaminate the machine and disconnect it from mains power and gas supplies before removing panels, and before return to use run the full leak test and self-test to confirm the cleaned flow sensors read correctly.
Lesson: A clogged inlet filter eventually lets dust through to the flow sensors, so when an anesthesia machine keeps drifting out of calibration, check and service the filter first.