Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / EKF Biosen C_line
EKF · Biosen C_line

EKF Biosen C_line: error at the first calibration step

What happened

As reported in the case, a glucose/lactate analyzer showed error 1105 at the first step of calibration, pointing to a suspected blockage in the fluidic path.

What was checked

Cause

Probable, as reported in the case: a clog in the fluidic path. The engineers found a blockage during disassembly and cleaning, and the error did not come back after the line was cleared.

What was done

The unit has no power switch, so full isolation means pulling the mains plug. Access to the internals is through a service opening in the base, with the needle head positioned first; the clog was removed and the nodes flushed and cleaned. The exact disassembly steps are a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Unplug at the wall before any disassembly — on this unit that is the only full isolation — treat residues in the line as biohazardous and wear gloves, and pass the electrical safety test before returning the analyzer to use.

Lesson: On analysers that must run continuously, know in advance how they are fully isolated (mains plug, not a button) and where the documented access points are, so a simple clog does not turn into long downtime.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes