Electron fluorograph: network link between operator and doctor workstations drops

What happened

Connection between the lab assistant workstation and the doctor workstation on the unit went down. The engineers reported that after a switch reboot a couple of packets got through, then requests started timing out.

What was checked

Cause

The engineers found an IP address conflict: a second Electron unit had been put on the same network with the same address, so the two devices clashed over it and the link dropped shortly after it came up.

The cause is probable and was not fully confirmed.

What was done

The engineers separated the addresses: each unit got a unique IP, and the sending workstation was pointed to the correct destination. The outcome was reported as probable, so treat the duplicate IP as the likely cause rather than fully confirmed.

Safety

Before unplugging or re-routing any cabling on an X-ray unit, isolate the system from mains so no exposure can be started while you work, and confirm the unit operates normally before returning it to use.

Lesson: When a network link on medical equipment drops right after coming up, check for a duplicate IP address on the same network before blaming the switch or the cable.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.