Electron fluorograph with EOS software: workstation freezes and slow database

What happened

The EOS software on the reporting workstation froze and lagged heavily, and every operation with the database was painfully slow.

What was checked

Cause

In this case the slowdown came from the Firebird 2.5 database server version: it performed poorly with this build of the EOS software, while the older 2.0 version ran normally.

What was done

The engineers rolled the Firebird server back from version 2.5 to 2.0, and the software response became effectively instant. Routine housekeeping was also noted as good practice: clearing out empty patient records and creating archive discs keeps the database compact and fast.

Safety

Shut down and unplug the workstation before any drive or software intervention, and before returning it to clinical use verify that image acquisition, saving and retrieval all work correctly.

Lesson: A sluggish workstation is not always a hardware problem — check the software stack, including database server version compatibility, before replacing parts.

Covered in the course: X-ray Service — Level 2: Fault-Finding →