Elmi SM-6MT: imbalance error at start after lid lock repair

What happened

Right after a lid lock repair, the centrifuge aborted every start-up with an imbalance error, even though the rotor itself was not the problem.

What was checked

Cause

The imbalance sensor calibration had been lost and its readings were sitting at zero, so the unit registered an imbalance on every run — probably disturbed during the earlier repair, though the case does not state the reason for certain.

What was done

The engineers restored the sensor's baseline reading to its correct value using a calibration procedure from the manufacturer's service documentation, done by authorised personnel. After calibration the unit started normally; the case was reported as solved.

Safety

Imbalance protection on a centrifuge guards against rotor failure at speed — never retune or silence it just to clear a fault, and before return to use verify the lid interlock and run a supervised test spin with a properly balanced rotor.

Lesson: When a new fault appears right after a repair, check the sensors and calibrations the work may have disturbed — a protection reading zero is something to restore by the documented procedure, not a setting to tune around.