Emotion 6 CT: previous studies fail to open on the review workstation (Pixel Data error)

What happened

On the physician's workstation, opening any previous study from the local database ended in an error related to Pixel Data.

What was checked

Cause

Not established in the case. The most likely explanation was corruption or inconsistency of the local study database on the review workstation: the studies were still listed, but their pixel data could no longer be retrieved.

The cause is probable and was not fully confirmed.

What was done

Reported fix, with a probable outcome: rebuild the local databases using the system's remove/recreate-all-databases batch from the utils folder, run under an administrator account; after a reboot the databases are recreated automatically. No host reinstallation was needed. Note that local studies not archived elsewhere are lost during the rebuild.

Safety

Take the scanner out of clinical use for the rebuild, make sure no examination is running or scheduled on it, and let it complete the full startup self-checks before it is returned to service.

Lesson: When a review workstation refuses to open every archived study, suspect the local database first — and rebuild it only after confirming the data is safely archived elsewhere.