Erba XL-640: needle stopped dispensing sample
What happened
The sampling needle stopped dispensing sample. No error codes were shown in the case.
What was checked
- Inspected the sampling needle and found it clogged; cleaned it.
- Exercised the syringes from the settings section to confirm fluid was moving.
- With the front panel removed, watched the syringes work to localize the blockage — pump, valve or needle line.
- Flushed the cuvettes with a diluted bleach-and-water solution fed through the system canister.
Cause
The engineers found the primary cause was a clogged sampling needle. They also suspected slime from the deionized-water canister that may have entered the syringes, so the whole fluid path was treated, not just the needle.
What was done
Cleaned the needle, ran the syringes through their stroke, and flushed the system with diluted bleach, including a cuvette wash. As reported in the case: if the syringes run free but dispensing still fails, check the pump and the valve line; the reported valve service life is 3–5 years. The analyzer was returned to use after rinsing.
Safety
Wear gloves and eye protection when preparing and running bleach solutions, and rinse the fluid path thoroughly with clean deionized water before returning the analyzer to samples.
Lesson: When a needle stops dispensing, treat it as a fluid-path blockage and localize it step by step from the needle inwards, while also checking the water canister as a possible contamination source.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.