Esaote Mylab Alpha: images swapped between the two displays
What happened
Both displays of the ultrasound system were working, but each showed the image meant for the other — the picture appeared on the wrong screen. No error codes were reported.
What was checked
- Confirmed that no external monitor was connected via HDMI and that the cabling was as expected.
- Looked for anything that could have changed the setup — in this case a plain USB keyboard was connected to the system, the easiest way to get into the BIOS.
- Entered the BIOS setup and checked the display initialization order.
- Watched at power-up which monitor shows the BIOS/boot screen — after a reset, that display becomes the primary one.
Cause
The engineers found the likely cause was a reset of the BIOS settings — apparently made via the plain USB keyboard connected to the system — which changed the order in which the monitors are initialized. This was treated as probable, not confirmed.
The cause is probable and was not fully confirmed.
What was done
Enter the BIOS setup and restore the correct monitor initialization sequence. Practical point from the case: after a BIOS reset, the display on which the BIOS screen appears at boot becomes the primary one, so set the order with that in mind. The outcome was reported as probable — the case was still being worked at the time.
Safety
Before returning the system to clinical use, run the full electrical safety test, including earth and patient leakage currents — BIOS and display work is still work on patient-connected equipment.
Lesson: A swapped picture on a multi-display system is often nothing more than a reset BIOS setting — check the display initialization order (and whether an ordinary USB keyboard is connected) before suspecting hardware.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.