Esaote MyLab: video chip overheats when the system drive fails
What happened
On aging MyLab platforms (boards M685, M742, M732 mentioned in the case), the custom graphics chip began overheating after the system hard drive failed: the operating system would not boot, the video drivers never started, and the chip heated up fast. The engineers noted that on some boards (M732) these chips seem to fail more often.
What was checked
- Blow out the ventilation openings and check that air actually flows through the chassis.
- Clean the cooling system regularly — animal hair builds up fast in veterinary use; clean or replace the blower if airflow is weak.
- Check that the copper heat-spreading plates on the graphics chip are fitted and making good contact.
- Monitor the health of the operating system and the system drive: if the drivers cannot start, the chip overheats.
Cause
As reported in the case: when the hard drive fails, the OS does not load and the video drivers never start, so the custom graphics chip runs uncontrolled and overheats. These chips are custom parts, not available on the market, so a failed one effectively cannot be replaced. The link between the drive failure and the chip damage was considered probable, not fully proven.
The cause is probable and was not fully confirmed.
What was done
No repair was possible for a dead chip; the reported work was prevention: keep the cooling clean, fit or renew the copper heat-spreading plates, and keep the OS and system drive healthy so the video drivers always start. The engineers also noted, on another platform in the family, separate problems with the power supply and the second probe connector — worth checking during preventive service.
Safety
Disconnect the unit from mains before opening the cabinet and let the internals cool: the graphics chip, heatsinks and PSU area can stay hot, and PSU capacitors can hold a charge — never touch components with power applied.
Lesson: On aging platforms with unobtainable custom video chips, a healthy operating system and clean cooling are the only real protection — prevention beats any repair.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.