Esaote · MyLab20

Esaote MyLab20: system software image fails to boot

What happened

A software image intended for a MyLab20 was put on the scanner, but the unit would not start with it. The engineers found no fault reports — the machine simply did not boot from that image.

What was checked

Cause

Despite the similar names, the MyLab20 and MyLab20+ are different machines: the plain 20 shares its platform and software with the MyLab15 generation, while the 20+ runs the same software family as the MyLab25/30/40/50 line. In this case the image on the drive did not match the firmware version of the ultrasound front-end.

The cause is probable and was not fully confirmed.

What was done

After confirming the exact model and the front-end firmware version, the engineers placed a system-software image from a compatible model of the matching platform onto the hard drive — the machine's identity is not defined by the image, the versions simply have to match. The reload itself was a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome is reported as probable: with matched versions the scanner was expected to start normally.

Safety

Before opening the scanner to service the drive, disconnect it from the mains and follow ESD precautions, and complete the electrical safety checks before the unit returns to clinical use.

Lesson: Check the exact model variant and front-end firmware version before imaging system software — similar model names can hide completely different platforms.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.