Esaote MylabSeven: operator panel touchscreen not responding

What happened

The touchscreen of the operator panel stopped responding to touch, while the ultrasound image itself kept displaying normally.

What was checked

Cause

In this case the engineers found contamination on the touch glass, most likely ultrasound gel or another liquid that had worked its way inside and blocked the touch sensing. This is a probable cause for this one case, not a general rule.

What was done

The screen was disassembled and the touch glass was cleaned with a dry lint-free cloth, then reassembled. The touchscreen returned to normal operation; no parts were replaced.

Safety

Disconnect the system from the mains and let internal capacitors discharge before opening the console, and after any work inside the unit run an electrical safety test before returning it to clinical use.

Lesson: When a touchscreen shows a picture but ignores touch, check the touch glass for gel or fluid contamination before suspecting the electronics.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.