Euronda E9: chamber fills with water after the cycle
What happened
After a completed sterilisation cycle and shutdown, the chamber filled with distilled water on its own, and water spilled out when the door was opened.
What was checked
- Checked the s11 valve and the calibration valve for tight closing.
- Checked the tank valves — the suspicion was that they were not shutting and water was feeding the chamber by gravity.
- Watched the water behaviour while the chamber cooled with the door open.
- Checked the check valve on the steam release line.
Cause
The engineers found the inlet valves (s11 and the calibration valve) were not closing, so water from the reservoir fed the chamber by gravity even with the unit off.
What was done
The faulty valves were replaced. Second-hand valves were fitted as a temporary measure and stopped the self-filling; new valves were ordered as the permanent fix.
Safety
Before working on the water circuit, isolate the unit from the mains and let the chamber, reservoir and steam paths cool, depressurise and drain fully, as residual hot water and steam can scald; after the repair, check for leaks and run a full cycle before returning the steriliser to use.
Lesson: Water slowly accumulating in the chamber with the unit off points first to inlet valve seats not holding — swap-test the valves before suspecting the check valve or the controls.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →