Ферропласт · АЭ-14-ЯФп

Ferroplast АЭ-14-ЯФп: intermittent dirty distillate

What happened

Unstable operation: the unit worked intermittently, and the distillate quality periodically fell out of spec — the water came through unclean at times.

What was checked

Cause

The reported cause was insufficient flow of running cooling water through the distiller: the steam did not condense fully, and hot steam carrying impurities passed into the distillate outlet — hence the intermittent dirty water.

What was done

The engineers moved the cooling water intake to a different source (another tap-off point). After that the cooling flow was stable and the distillate quality returned to normal. The case was reported as solved.

Safety

A distiller holds boiling water and live steam — shut it down, isolate the power and water supply, and let it cool completely before working on the cooling circuit; confirm the distillate quality by analysis before returning the unit to use.

Lesson: Intermittent distillate faults usually start upstream of the unit — before blaming the still, verify the cooling water flow and the stability of its source.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.