Ferroplast АЭ-14-ЯФп: intermittent dirty distillate
What happened
Unstable operation: the unit worked intermittently, and the distillate quality periodically fell out of spec — the water came through unclean at times.
What was checked
- Checked the quality of the incoming feed water against the norm.
- Monitored the flow of cooling water through the distiller's cooling circuit.
- Analysed the water at the source, at the intake point.
Cause
The reported cause was insufficient flow of running cooling water through the distiller: the steam did not condense fully, and hot steam carrying impurities passed into the distillate outlet — hence the intermittent dirty water.
What was done
The engineers moved the cooling water intake to a different source (another tap-off point). After that the cooling flow was stable and the distillate quality returned to normal. The case was reported as solved.
Safety
A distiller holds boiling water and live steam — shut it down, isolate the power and water supply, and let it cool completely before working on the cooling circuit; confirm the distillate quality by analysis before returning the unit to use.
Lesson: Intermittent distillate faults usually start upstream of the unit — before blaming the still, verify the cooling water flow and the stability of its source.