Flex Endo 1: water draining into the disinfectant container

What happened

A washer-disinfector sometimes drains water into the container holding the disinfectant solution instead of the intended drain path.

What was checked

Cause

Probable cause reported in the case: one of the electronic drain valves no longer closes tight and lets water through into the wrong path — here, into the disinfectant container. A similar case mentioned the sewer-side valve not closing during the disinfectant drain, so either of the two valves can be the culprit; confirm before replacing.

The cause is probable and was not fully confirmed.

What was done

No confirmed repair in the case. The recommendation was to test both drain valves for leak-through in the closed position and replace the one that fails; after replacement, run a full cycle and verify that every phase drains where it should. The outcome in this case was probable, not confirmed.

Safety

Treat the disinfectant container, drain lines and any spillage as chemically active and contaminated: wear PPE, disconnect and lock out power before working on the valves, and verify a full disinfection cycle before returning the machine to use.

Lesson: When water ends up in the wrong container, suspect a drain valve that no longer shuts tight and test every valve on the drain paths before replacing anything.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.