Fujifilm DryPix Smart: films printing too light
What happened
The imager started printing films noticeably too light (under-exposed). Loading a new batch of film did not help.
What was checked
- Printed a test/calibration image to confirm the density fault was systematic, not a one-off
- Checked the printer's web interface and what film data the unit had read at the last pack change
Cause
The engineers found that when the new film pack was loaded, the printer misread the pack's film data and applied a wrong calibration; the hardware itself was healthy. It was also suggested in the case that an old backing sheet left in the magazine at loading can make the pack tag unreadable.
What was done
The film data reading fault was cleared with a procedure from the manufacturer's service documentation, done by authorised personnel. After that the printer returned to normal density; no parts were replaced.
Safety
Before returning the imager to clinical use, run the manufacturer's density/calibration check and confirm the printed films meet acceptance criteria, since a mis-calibrated imager degrades diagnostic image quality; disconnect mains power before reaching into the film path.
Lesson: When print density changes suddenly, check how the last film pack was loaded and what the printer actually read from it before suspecting the hardware.