Fujinon · 4450

Fujinon 4450: xenon lamp hour counter due for reset

What happened

On a Fujinon 4450 endoscopy system the light source kept reminding the engineers that the xenon lamp had reached its rated service life and that its hour counter needed to be reset. No fault codes were shown; the reminder itself was the whole complaint.

What was checked

Cause

The light source counts accumulated lamp hours and asks for a reset once the counter reaches the manufacturer's limit — normally the point where the lamp should be replaced. In this case no component had failed, as far as reported; the counter had simply done its job.

What was done

The engineers verified the actual condition of the lamp first, and the counter was then cleared as a procedure from the manufacturer's service documentation, done by authorised personnel. After the reset the counter read correctly, the lamp gave stable output, and the reminder did not return — the case was reported as solved.

Safety

A xenon lamp works under high pressure and emits intense UV: let it cool, isolate the unit from mains before opening the lamp compartment, use eye protection and clean gloves when handling the bulb, and run the electrical safety check plus a light-output check before the source returns to clinical use.

Lesson: A lamp-hour counter is a protection, not a nuisance — reset it only as part of the documented lamp replacement procedure, never simply to clear the reminder.

Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.