Fujinon EG-530 / EC-530WL3: water inside the scope during washing despite a passing leak test
What happened
The engineers found that video gastroscopes EG-530 and colonoscopes EC-530WL3 reported as leak-tight still took in water during the washing stage of reprocessing.
What was checked
- Review the whole leak-test routine as actually performed: test pressure, immersion time, and what is done with the scope under water.
- Check whether the leak tester is disconnected while the scope is still submerged.
- Test the connector itself: pressurise the scope, remove the cap, submerge the connector end in water and watch for bubbles.
- Inspect the tubing between the leak tester and the scope, and the venting path, for trapped water.
Cause
Probable cause: water sitting in the tubing between the leak tester and the endoscope. In this case the connector had been uncapped and left in the washing tray, so water could reach the connector and be pushed inside during the next test or wash.
The cause is probable and was not fully confirmed.
What was done
No completed repair is recorded in the case. The agreed corrective practice was: never leave the uncapped connector standing in the tray, drain and dry the tester tubing before attaching it, then repeat the leak test — including the connector check — before the scopes go back into the reprocessing loop.
Safety
Endoscopes are critical infection-control items: if water ingress is suspected, take the scope out of service, fully reprocess it, and repeat a documented leak test before it is used on a patient.
Lesson: A passing leak test only certifies the moment of the test — water can still enter through wet tester tubing or an uncapped connector left in the tray.
Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.