Furuno CA-270: sample tubes flagged red with no errors
What happened
The analyzer highlighted sample tubes in red while showing no error codes, and the reported results were within their normal ranges.
What was checked
- Clean and polish the sampler needle tip with a fine abrasive (grit 2500–3000, as reported in the case).
- Check the condition and freshness of the QC material — it may have been reconstituted or diluted long before use.
- Review the calibration history and calibration factors for signs of a calibration run on depleted reagent.
Cause
The cause was not finally confirmed in this case. The suspicion, reported in the case, was stale or long-diluted QC material, or a calibration performed on depleted reagent that produced high calibration factors — enough to flag tubes without raising any error.
The cause is probable and was not fully confirmed.
What was done
The engineers recommended removing and washing the sampler with a hypochlorite (bleach) solution, polishing the needle tip, and verifying QC freshness together with the calibration history. The outcome was reported as probable, so treat this as a likely direction rather than a confirmed fix.
Safety
Treat the sampler and tubes as potentially infectious, use the hypochlorite concentration from the manufacturer's documentation and rinse thoroughly to protect the needle from corrosion, and confirm QC passes before returning the analyzer to patient testing.
Lesson: When samples get flagged without an error code, check the reference side of the chain — fresh QC material and a calibration done on fresh reagent — before suspecting hardware.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.