GE (General Electric) · 1.0Т

GE 1.0T MRI: gradient block temperature alarm only at cold start-up

What happened

The fault appeared only at full power-up after the system had been sitting cold — reported in the case at around 20 °C on the units, typically the first start of the day. The system flagged the gradient block temperature sensor, but the reading was false: once warmed up, the system started with no complaint.

What was checked

Cause

A weakened solder joint on a board inside the gradient block. At cold start the joint lost contact, so the temperature sensor circuit gave a false alarm; after warm-up the contact returned and no fault appeared.

What was done

The engineers repaired the weakened solder joint on the board. After that the cold-start fault disappeared and the system returned to normal operation. A spare gradient block was available as backup, but it was not needed.

Safety

An MRI magnet stays energised at all times: keep ferromagnetic tools and parts well clear of the scan room during such work, and complete the manufacturer's electrical safety checks before returning the system to use.

Lesson: A fault that appears only at cold start usually points to a temperature-sensitive connection such as a weakened solder joint, so chase the boards and connections before condemning the sensor.