GE · Acuson (Акувикс)

GE Acuson (Acuvix): no image on monitor or touch panel at power-on

What happened

The unit powers on normally, but there is no picture on either the main monitor or the touch panel.

What was checked

Cause

One of the RAM modules in the image PC had failed, so the system started but produced no video output to either display.

What was done

Replaced the faulty RAM module. The engineers found that image output to the monitor and touch panel was restored after the swap, and the video card, CPU and motherboard were confirmed serviceable.

Safety

Disconnect the system from mains power and take ESD precautions before handling memory modules or boards, and repeat the electrical safety test before returning the unit to clinical use.

Lesson: When the console powers up but both screens stay dark, troubleshoot the image PC first like any ordinary computer — boot, video path, memory — before suspecting the ultrasound chain itself.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.