GE Avance 5: breathing bag inflates overnight on a powered-off machine
What happened
Reported in the case: the reservoir breathing bag inflated on its own overnight while the anesthesia machine was switched off.
What was checked
- The engineers traced the leak path by switching ventilation between mechanical and automatic modes.
- The rubber sealing membrane in the mode-switching block was examined, and the engineers found cracks in it.
Cause
The probable cause reported in the case: cracks in the rubber membrane that plugs the port when switching between mechanical and automatic ventilation let gas slowly leak into the bag circuit even with the machine off.
The cause is probable and was not fully confirmed.
What was done
The reported fix was partial disassembly of the switching block and replacement of the rubber membrane. Note that the cause was rated probable in the case, not fully confirmed.
Safety
Before opening the gas block, shut off and disconnect all gas supplies and relieve residual line pressure — a sealed anesthesia circuit can store enough gas pressure to eject parts or vent gas during disassembly.
Lesson: A powered-off machine can still self-pressurize: slow leaks in mode-switching seals show up overnight, so inspect the rubber membranes in the switching block whenever a reservoir bag creeps up on a shut-down unit.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.