GE AW 4.7: screen resolution resets after reboot or sleep
What happened
On a GE AW 4.7 workstation the screen resolution could not be changed and kept: any setting made through the standard display tools was lost after a reboot or after sleep mode.
What was checked
- Tried to set the resolution with the workstation's own display settings — it applied but would not stick.
- Rebooted the station and woke it from sleep; the setting was reset every time.
- Looked at the X server (xorg) configuration that defines the display mode at startup — it is set up for the approved monitors.
Cause
In this case the X server rebuilt its display configuration at every start from its own file, tuned to the approved monitors, so anything chosen through the GUI was simply overwritten.
What was done
The engineers pinned the required display mode by editing the X server (xorg) configuration on the workstation; after that the resolution reported in the case (1920×1080) held across reboot and sleep. Keep a backup of the original configuration file before any such edit.
Safety
Before returning the station to use, verify image geometry and test patterns against the display requirements and confirm the workstation self-check passes — a wrong display mode on a review station can distort what the clinician sees.
Lesson: On locked-down OEM workstations the display mode usually lives in the X server configuration rather than the GUI — pin it there, back up the original file, and verify the image before the unit goes back to work.
Covered in the course: DICOM, PACS and Networks — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.