GE Brilliance: gantry rotor brake sticking with burning smell
What happened
As reported in the case, the gantry rotor brake seizes and a strong burning smell comes from the brake; the engineers said this complaint is frequent on this model, but that is one case, not a rule.
What was checked
- De-energise and lock out the gantry, let the brake cool, then inspect it for heat discolouration, burnt lining smell and signs that it never fully releases and drags on the rotor.
- Check brake release and engagement against the manufacturer's documentation, so that braking by the drive is not masking a brake that sticks.
- After any brake work, test that emergency-stop braking of the gantry actually works before the CT goes back to clinical use.
Cause
The gantry rotor brake sticking in a dragging position and overheating — the burning smell reported in the case points to the coil and friction lining running hot; the engineers treated it as a known weak spot of the model, but that is their experience in this one case.
What was done
The case was reported solved, but the workarounds discussed there rely on taking the emergency-stop brake out of the circuit, so they are not something to copy: the acceptable fix stands as a procedure from the manufacturer's service documentation, done by authorised personnel, with the brake left fully functional and E-stop braking verified before return to use.
Safety
De-energise and lock out the gantry before touching the brake — the rotor carries stored kinetic energy and can coast, the assembly is hot enough to burn, and the CT must not return to use until emergency-stop braking has been proven.
Lesson: A safety-related brake that sticks and smells hot must be repaired, not neutralised — workarounds in the emergency-stop chain turn a small fault into an incident.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.