GE · Brivo MR355

GE Brivo MR355: table drive motor seized

What happened

The patient table stopped moving. The drive motor was seized and would not turn. No error codes were reported in the case.

What was checked

Cause

Mechanical failure of the table drive motor: a piece of the stator magnet broke off and jammed the mechanism.

What was done

The motor was opened, the detached magnet fragment and debris were cleaned out, and the unit worked again — the outcome was solved in this case. A replacement motor was being sought for spares, since a motor with a damaged magnet is not trusted for long-term service.

Safety

De-energise the drive before opening it, and respect the MRI static field: a detached magnet fragment and ferromagnetic tools near the bore can become projectiles — check the manual table release, the emergency stop, and complete an electrical safety test before the table returns to use.

Lesson: A seized table drive is not automatically the gearbox — open the suspect component and look inside before ordering parts.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.