GE Carestation 650: rapid sevoflurane consumption

What happened

Sevoflurane was being consumed unusually fast; the vaporizer itself had already been swapped and the high usage remained unexplained.

What was checked

Cause

The engineers found the seat O-rings on the vaporizer mounting station worn: they had changed shape and lost elasticity, so sevoflurane was evaporating out around the seat instead of staying sealed.

The cause is probable and was not fully confirmed.

What was done

Moved the vaporizer to the second mounting station (reported in the case as the spare, with intact O-rings); in effect the worn O-rings are the part to replace. The outcome of the fix was reported as probable, not fully confirmed.

Safety

Before the machine returns to clinical use, run the manufacturer's leak and agent-delivery checks, and remember that escaping anaesthetic vapour is an inhalation hazard — ventilate the work area or use gas scavenging.

Lesson: When anesthetic consumption runs high, do not stop at the vaporizer itself — the mounting-station seals are a classic silent leak path and are cheap to inspect first.