GE Carestation 650: rapid sevoflurane consumption
What happened
Sevoflurane was being consumed unusually fast; the vaporizer itself had already been swapped and the high usage remained unexplained.
What was checked
- Swapped the sevoflurane vaporizer for another unit to rule out the vaporizer itself.
- Inspected the vaporizer mounting station and its seat O-rings for flattening, hardening and loss of elasticity.
Cause
The engineers found the seat O-rings on the vaporizer mounting station worn: they had changed shape and lost elasticity, so sevoflurane was evaporating out around the seat instead of staying sealed.
The cause is probable and was not fully confirmed.
What was done
Moved the vaporizer to the second mounting station (reported in the case as the spare, with intact O-rings); in effect the worn O-rings are the part to replace. The outcome of the fix was reported as probable, not fully confirmed.
Safety
Before the machine returns to clinical use, run the manufacturer's leak and agent-delivery checks, and remember that escaping anaesthetic vapour is an inhalation hazard — ventilate the work area or use gas scavenging.
Lesson: When anesthetic consumption runs high, do not stop at the vaporizer itself — the mounting-station seals are a classic silent leak path and are cheap to inspect first.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.