GE claude-haiku-5-5: patient data entered with errors after software hang

What happened

After the console software froze, patient data had to be typed in manually and was saved with mistakes. The engineers needed to correct the already-created study record on the unit itself.

What was checked

Cause

The software hang forced manual entry of patient data, and that manual entry contained mistakes. The root cause of the hang itself was not diagnosed in the case.

What was done

On the unit, open the affected patient, open the Edit tab on the right, confirm, and correct the data. When prompted, enter the requested operator identifier (surname or a short three-letter signature, as reported in the case) to confirm the edit, then confirm again and send the corrected study to the server manually. Per the case, editing only became possible after another study had been performed on the unit. Finally, delete the flawed study from every host it had already been sent to and manually resend the corrected study.

Safety

After any software hang, restart the system only per the manufacturer's documentation and make sure it runs stably before accepting the next patient, and always check corrected identity data against the patient's records before the study is used or sent — a mislabelled study can lead to a wrong-patient decision.

Lesson: Verify manually entered patient data at the moment of entry — especially after a software hang — because correcting a study afterwards is slower and leaves flawed copies to hunt down on every connected host.