GE CTe: aging X-ray tube checked before Qcal calibration

What happened

Before a planned Qcal calibration there was doubt whether the X-ray tube (reported in the case as 3.5 years in service with low workload) would withstand it; an earlier software installation had also been blocked by a corrupted save state.

What was checked

Cause

No actual fault was found. The concern was calibration stress on an older, lightly used tube; the pre-checks revealed no rotor noise and no arcing, so nothing pointed to worn-out tube condition.

What was done

With the pre-checks clean, the Qcal calibration was performed according to the manufacturer's service documentation and completed successfully — the case is considered solved.

Safety

Before pulling or reseating high-voltage receptacles, power the system down and allow the HV cables and circuits to fully discharge — stored charge in the tube, cables and capacitors can be lethal.

Lesson: Before putting calibration stress on an aging tube, confirm it can take it — rotor noise, arcing and clean, greased HV receptacles are the quick checks that decide.