GE Discovery 710 PET/CT: CT series cannot be confirmed after warning

What happened

When the study moved to the CT stage, a warning window appeared and the confirm button stayed inactive, so the scan could not continue.

What was checked

Cause

The engineers found the protocol requested too many slices for one series; the system refused confirmation, apparently to protect the tube from overheating. Wrong positioning or scan coverage can trigger the same warning.

What was done

Disable the group-by-group (batch) acquisition mode in the protocol. If the warning persists, reduce the number of slices by increasing the slice interval, or narrow the scan field/coverage, then re-check the protocol.

Safety

Respect the X-ray tube thermal capacity: stacking too many acquisitions in one series risks tube overheating and gives the patient avoidable radiation dose.

Lesson: A greyed-out confirm button at the CT handoff is usually the scanner protecting itself — check the protocol's slice count and coverage before suspecting hardware.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.