GE Engström: suspected poor air/oxygen gas delivery

What happened

In this case the machine was suspected of incorrect gas delivery through the air/oxygen supply valves, and the engineers also discussed whether valves with different part numbers are interchangeable.

What was checked

Cause

As reported in the case: in a similar earlier case the valve was blamed during calibration, but the real cause was the reducer, which was not delivering normal flow — an upstream restriction looks just like a faulty delivery valve.

The cause is probable and was not fully confirmed.

What was done

The engineers advised checking the reducer and the inlet filters before condemning the valves. As reported in the case, a valve marked 2100575 was fitted in place of 1505-3209-000-S, passed calibration and worked normally; the outcome is probable, not confirmed.

Safety

Before opening the pneumatic block, isolate the ventilator from mains power and gas supplies and bleed the residual pressure — compressed gas is stored energy — and after any valve or regulator work, complete the manufacturer's full leak and delivery verification before the unit is returned to a patient.

Lesson: Before replacing a gas delivery valve, prove the whole supply path — filters, pressure and the reducer's flow capacity — because an upstream restriction is the classic false culprit.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.