Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / GE Engstrom
GE · Engstrom

GE Engstrom: wrong displayed volume while delivered volume looks normal

What happened

The ventilator showed incorrect volume readings although the lungs inflated normally and actual delivered volume appeared unchanged; replacing the flow sensor did not correct the display.

What was checked

Cause

No confirmed cause was recorded in the case. The engineers' working suspicion was a measurement problem in the breathing block rather than a real change in ventilation — most likely an internal leak or the flow-measurement chain (exhalation valve membrane seating, sensor calibration/drift), since the external flow sensor alone had already been swapped without effect. This remains probable, not confirmed.

The cause is probable and was not fully confirmed.

What was done

The final fix was not confirmed in the case. The agreed path was to restore and verify tightness of the breathing block and exhalation valve, then recalibrate the flow sensors and confirm displayed volumes against a test bag; calibration and teardown steps, where restricted, are a procedure from the manufacturer's service documentation, done by authorised personnel. Outcome was reported as probable.

Safety

Treat the breathing block and circuit parts as biohazard-contaminated during teardown, and before return to use verify displayed volumes against a calibrated test lung and run the manufacturer's full pre-use check, because a ventilator showing wrong volumes can silently under- or over-ventilate a patient.

Lesson: When the displayed volume disagrees with what the lung visibly receives, work through the whole measuring path — valve, seals, calibration — and confirm on a test lung before trusting the display.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes