GE · HDxt

GE HDxt: slow helium loss after cold head replacement

What happened

Some time after a cold head replacement, the magnet was losing helium at about 1% per day as reported in the case. Cryostat pressure held at its usual value (around 1 psi reported in the case), and there was no frosting anywhere on the system.

What was checked

Cause

The engineers found a microcrack in one of the silicone O-ring seals in the helium circuit. The leak was small enough that pressure still held and nothing frosted externally — likely disturbed or damaged during the earlier cold head work.

What was done

The engineers replaced the cracked silicone O-ring with a new seal; the reported leak stopped and the helium level stabilised.

Safety

Treat the cryostat as a pressurised cryogenic vessel: ensure ventilation and oxygen monitoring in the magnet room, avoid contact with cold surfaces and escaping gas, and never obstruct, plug or defeat relief devices such as the burst disc — if helium loss suddenly accelerates, treat it as a quench risk and hand over to authorised personnel.

Lesson: A slow helium leak with normal pressure and no frosting points to small seals — chase O-rings and connections with a leak detector before suspecting the cold head or the vessel itself.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.