GE HDxt: slow helium loss after cold head replacement
What happened
Some time after a cold head replacement, the magnet was losing helium at about 1% per day as reported in the case. Cryostat pressure held at its usual value (around 1 psi reported in the case), and there was no frosting anywhere on the system.
What was checked
- Soap-tested the burst disc and the relief valve; no bubbles appeared.
- Checked cryostat pressure and inspected the system for icing; both looked normal.
- Monitored the cold head heater duty cycle via the magnet monitor to see whether the load pointed to losses in the helium circuit.
- Swept the cryogenic connections with a helium leak detector to localise the loss.
- Inspected the burst disc for cracks.
Cause
The engineers found a microcrack in one of the silicone O-ring seals in the helium circuit. The leak was small enough that pressure still held and nothing frosted externally — likely disturbed or damaged during the earlier cold head work.
What was done
The engineers replaced the cracked silicone O-ring with a new seal; the reported leak stopped and the helium level stabilised.
Safety
Treat the cryostat as a pressurised cryogenic vessel: ensure ventilation and oxygen monitoring in the magnet room, avoid contact with cold surfaces and escaping gas, and never obstruct, plug or defeat relief devices such as the burst disc — if helium loss suddenly accelerates, treat it as a quench risk and hand over to authorised personnel.
Lesson: A slow helium leak with normal pressure and no frosting points to small seals — chase O-rings and connections with a leak detector before suspecting the cold head or the vessel itself.