GE HDxt: workstation hangs polling USB and won't boot after software reinstall

What happened

After a standard software reinstallation on the workstation (OS layer, then application) and a reboot, the system failed to start and appeared stuck in an endless cycle of polling USB devices.

What was checked

Cause

In this case the engineers found three mice connected to the workstation at once; during startup the software could not select one of them and the boot kept looping on USB enumeration.

What was done

The extra mice were disconnected, leaving a single pointing device, and the system then booted normally. The case was reported as solved.

Safety

Shut the workstation down properly and remove mains power before connecting or disconnecting peripherals; after any service work, complete the electrical safety checks before the scanner is returned to clinical use.

Lesson: Before blaming a software installation, rule out the simplest hardware conflicts — several identical USB input devices on one port chain can stall a boot on their own.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.