GE Innova 3100: table rises with delay and abnormal noise
What happened
Reported in the case: the table began to rise with a delay and an abnormal noise, while on a cold start everything worked normally; sparking was seen at a relay.
What was checked
- Joystick checked for liquid ingress or contamination — found serviceable.
- Motor brake inspected; the gap was found out of adjustment.
- Relay examined; sparking was observed when the drive was switched.
- Cold-versus-warm behaviour noted, pointing away from electronics and toward mechanical drag.
Cause
The engineers found an excessive gap in the motor brake: the brake loaded the motor, so the drive started with a delay and noise, and the relay sparking was a secondary sign of the heavier motor current (as reported in the case).
What was done
The brake was cleaned and its gap re-adjusted per the manufacturer's service documentation. After that the table moved without delay or noise, and the relay sparking stopped. Case closed as solved.
Safety
Before opening the vertical drive, isolate the system and mechanically secure the tabletop — with the brake released it can drift or drop under its own weight — and after any brake work confirm the brake holds the table before returning it to use.
Lesson: A motion fault that appears only when warm can come from mechanical drag such as an out-of-adjustment motor brake, so check the brake before blaming joysticks or relays.