What happened

Reported in the case: the table began to rise with a delay and an abnormal noise, while on a cold start everything worked normally; sparking was seen at a relay.

What was checked

Cause

The engineers found an excessive gap in the motor brake: the brake loaded the motor, so the drive started with a delay and noise, and the relay sparking was a secondary sign of the heavier motor current (as reported in the case).

What was done

The brake was cleaned and its gap re-adjusted per the manufacturer's service documentation. After that the table moved without delay or noise, and the relay sparking stopped. Case closed as solved.

Safety

Before opening the vertical drive, isolate the system and mechanically secure the tabletop — with the brake released it can drift or drop under its own weight — and after any brake work confirm the brake holds the table before returning it to use.

Lesson: A motion fault that appears only when warm can come from mechanical drag such as an out-of-adjustment motor brake, so check the brake before blaming joysticks or relays.