Academy / Field fault notes / GE Logiq 5 Pro
GE · Logiq 5 Pro

GE Logiq 5 Pro: starts on its own and no image

What happened

As soon as power is applied, the system unit starts by itself and does not respond to the power button (the button itself is good). No image on the built-in or an external monitor. With the video capture board cable disconnected, the external monitor shows a signal and the system boots, but then reboots repeatedly.

What was checked

Cause

Not confirmed — the outcome in this case was only probable. The engineers' working hypothesis: with no supply on the video capture board and the ultrasound front-end, the motherboard's power-on signal pin sits pulled to ground, so the board starts as soon as mains is applied instead of waiting for the button.

The cause is probable and was not fully confirmed.

What was done

Diagnosis was still in progress, so the fix is a direction, not a completed repair: trace why the ultrasound front-end does not power up and where the supply to the video capture board is lost, rather than fighting the button behaviour itself. Board-level work should follow the manufacturer's service documentation and be done by authorised personnel.

Safety

Disconnect the unit from mains before opening it and treat power supply capacitors as charged — they can hold a dangerous stored charge — and run the electrical safety test before the machine returns to clinical use.

Lesson: When a PC-based medical system powers up on its own and only works with an option board disconnected, hunt for the missing supply on that board first — the start-up logic is usually a symptom, not the fault.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.