GE Logiq 7: no digital board response, demo mode only

What happened

On diagnostic start-up the system reported it could not get digital board information. The probes made only a faint squeak and the image worked in demo mode, meaning there was no link between the ultrasound front end and the host PC.

What was checked

Cause

Suspected fault of the BF board (DDBF128) or its connection to the host PC, leaving the digital front end unreachable; the cause was probable, not confirmed in the case.

The cause is probable and was not fully confirmed.

What was done

The engineers advised to inspect and reseat the cable and connectors between the ultrasound (beamformer) side and the computer, and if the link stays down to treat the DDBF128 BF board as the likely part to replace. The outcome was probable at the time of the case.

Safety

Unplug the system from the mains and let the power supply capacitors discharge before opening the console, and complete an electrical safety check (protective earth and leakage currents) before the unit returns to clinical use.

Lesson: When a fault looks identical on every probe and even with probes unplugged, look at the shared digital front end and its link to the host, not at the transducers.