Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / GE Logiq A5 / Logiq P5
GE · Logiq A5 / Logiq P5

GE Logiq A5/P5: Incorrect operation after software install — nameplate does not match the hardware

What happened

After a software installation on this unit, the system worked incorrectly, and in the reported experience some software builds refused to run at all, reporting a hardware incompatibility. The installed software model simply did not match the machine's actual platform.

What was checked

Cause

A hardware-to-label mismatch: the case was badged Logiq A5, but the internal platform was from a Logiq P5, so the wrong model was picked during the software installation.

What was done

Software matching the actual platform — identified from the boot-time platform information — was installed instead of software matching the nameplate; the installation media is common to both models, so the model must be selected to match the real hardware. The reinstallation itself was a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Isolate the system from the mains before opening it to inspect the hardware, and after any software reinstallation complete the electrical safety test and confirm normal operation before returning the unit to clinical use.

Lesson: Verify the real platform from the hardware itself before choosing which software model to install — the nameplate on the case can lie.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes