Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / GE Logiq Book XP
GE · Logiq Book XP

GE Logiq Book XP: black screen, no boot after hard drive replacement

What happened

The internal hard drive failed on this portable ultrasound. After a new drive was fitted and the software was loaded onto it, the unit would not start: the screen stayed black at power-on.

What was checked

Cause

The boot loader on the new drive was missing or broken: the software files were on the disk, but the boot chain that hands control to Windows was not, so the machine never reached the operating system.

The cause is probable and was not fully confirmed.

What was done

In this case the engineers rebuilt the boot chain: a generic bootloader (grub2dos) was written to the new drive and configured to hand over to the Windows boot loader; the outcome was reported as probable. Full restoration of the system and application software on this scanner is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Before opening the unit, disconnect it from mains and remove the battery, and carry out the full electrical safety test (protective earth and leakage currents) before returning the scanner to clinical use.

Lesson: After a hard drive replacement, 'black screen, no boot' usually points to the boot loader rather than the software — rebuild the boot chain first before suspecting anything else.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes