GE Logiq Book XP: black screen, no boot after hard drive replacement
What happened
The internal hard drive failed on this portable ultrasound. After a new drive was fitted and the software was loaded onto it, the unit would not start: the screen stayed black at power-on.
What was checked
- The engineers reinstalled the system and application software on the new drive — the symptom did not change.
- They created a temp folder in the root of the system drive, as the application expects it — no effect on boot.
- They then checked the boot area of the new drive and found the Windows boot loader was not properly in place, so the unit had nothing valid to load at startup and stopped at a black screen.
Cause
The boot loader on the new drive was missing or broken: the software files were on the disk, but the boot chain that hands control to Windows was not, so the machine never reached the operating system.
The cause is probable and was not fully confirmed.
What was done
In this case the engineers rebuilt the boot chain: a generic bootloader (grub2dos) was written to the new drive and configured to hand over to the Windows boot loader; the outcome was reported as probable. Full restoration of the system and application software on this scanner is a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Before opening the unit, disconnect it from mains and remove the battery, and carry out the full electrical safety test (protective earth and leakage currents) before returning the scanner to clinical use.
Lesson: After a hard drive replacement, 'black screen, no boot' usually points to the boot loader rather than the software — rebuild the boot chain first before suspecting anything else.