GE Logiq: boot stops at bootloader after hard drive replacement
What happened
After a hard drive was replaced and a system image was written to the new drive, boot stopped at the bootloader initialisation stage. The engineers tried several different images and the hang was the same with each one.
What was checked
- Wrote several different system images to the new hard drive — boot stopped at the same bootloader stage every time
- Ruled out one corrupt image file, since no image got past the bootloader stage
Cause
The working conclusion in this case was that a cloned image taken from the old drive does not reliably boot on a Logiq workstation once deployed to a new hard drive — the image, not the new drive, was the likely problem. This is a probable, case-specific conclusion, not a rule.
The cause is probable and was not fully confirmed.
What was done
The recommendation was to stop using cloned images: obtain the original installation media and perform a clean installation of the system software — a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable.
Safety
Before opening the workstation, disconnect the unit from mains power and allow internal supplies to discharge, and after any software reinstallation complete the manufacturer's functional checks and an electrical safety test before the system returns to clinical use.
Lesson: An image cloned from the old drive cannot be assumed to boot on a new one, so before starting a system drive replacement make sure the original installation media and a clean-install procedure are available.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.