GE Logiq: convex probe loads the wrong preset

What happened

When selecting the convex probe, the system loaded the preset of a vaginal (endocavitary) probe. The convex-probe presets also dropped out from time to time.

What was checked

Cause

The engineers found the preset data in the software to be corrupted (probable cause). It was also noted in the case that preset dropouts on the convex probe had been seen more than once on this platform — that is one engineer's observation, not a general rule.

What was done

The fault was cleared only by a full reinstallation of the system software — a procedure from the manufacturer's service documentation, done by authorised personnel. A reboot alone resolved it only temporarily where it worked at all. After such a reinstall, presets and user settings should be verified and re-entered if needed.

Safety

After any software reinstallation, complete the manufacturer's verification steps and run an electrical safety test, including patient leakage current with all probes connected, before the machine returns to clinical use.

Lesson: When presets start mixing up or disappearing, a reboot is worth trying first, but stubborn preset corruption points to damaged software and only a proper reinstall to the manufacturer's documentation fixes it.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.