GE Logiq E: No boot after software image restore onto a smaller drive
What happened
After the Logiq E software image was deployed onto a 320 GB drive instead of the original 500 GB one (capacities as reported in the case), the system failed to boot and stopped with a GRUB loader error.
What was checked
- Check the capacity of the source drive the image was captured from — an image taken from a reported 500 GB drive will not deploy onto a smaller one.
- Confirm the software version on the unit and whether a proper installation image exists for it, rather than relying on a drive clone.
- Verify that the replacement drive meets or exceeds the original capacity before starting the restore.
Cause
The image had been captured from a larger (reported 500 GB) drive, and its partition layout physically cannot fit on a smaller 320 GB disk, so the boot loader had nothing valid to load. The engineers concluded the undersized target drive was the cause — the system simply expects the original capacity or larger, not a corrupted image.
What was done
Fit a drive of the original or larger capacity (500 GB in this case) and re-deploy the image; the engineers recommended using a dedicated disk-imaging utility for both capture and restore, and checking the software version in case a matching installation image is available. The system reportedly booted normally once the correct-capacity drive was installed.
Safety
Before opening the unit to swap the drive, power it down and disconnect from mains, take ESD precautions, and complete the electrical safety test before returning the system to clinical use.
Lesson: A drive image captured on a larger disk will not restore onto a smaller one, so always match or exceed the original drive capacity before cloning.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.