Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / GE LOGIQ E10
GE · LOGIQ E10

GE LOGIQ E10: boots into demo mode after CRX board swap

What happened

After the CRX3 receive-path board was replaced with a CRX2, the system started up in demo mode with an error message instead of normal scanning.

What was checked

Cause

Probable cause in this case: after the swap the system software/configuration no longer matched the installed hardware. An incompatible or faulty replacement board could give the same picture, and the case did not finally confirm which of the two it was.

What was done

Reported as resolved after a full software installation from scratch, with configuration done as a procedure from the manufacturer's service documentation by authorised personnel. The engineers also advised verifying the replacement CRX board itself; if the fault returns, fit a repaired or confirmed-good board.

Safety

Before replacing an internal board, power down, disconnect the unit from mains and allow internal capacitors to discharge, and complete an electrical safety test before returning the scanner to clinical use.

Lesson: Swapping a board to a different revision can leave the software configuration mismatched, so verify hardware compatibility first and be ready for a documented software reinstall rather than a simple board exchange.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes