GE Logiq e9: touch screen with dead zones and intermittent response
What happened
The touch screen registered presses only intermittently on the left and at the bottom of the display, the right side did not respond at all, and the lower part was completely dead.
What was checked
- Touch calibration was performed and verified through the manufacturer's service utilities; the engineers found the dead zones stayed exactly the same.
- The console was opened and the gap between the housing and the touch panel inspected for debris or pinched cabling; nothing abnormal was found.
Cause
No confirmed root cause was recorded. Since calibration gave nothing and no debris or mechanical interference was found, the engineers judged the touch panel itself to have failed — a stable zonal loss of response fits a degraded digitiser. This remained a probable conclusion in this single case.
The cause is probable and was not fully confirmed.
What was done
The fix agreed in the case was replacement of the touch panel; the part was ordered, so the repair had not been verified at the time and the outcome was probable.
Safety
Isolate the system from the mains and allow internal supplies to discharge before opening the console, and repeat the electrical safety tests (protective earth and touch/leakage current) after reassembly before the unit returns to clinical use.
Lesson: Persistent dead zones that survive calibration and a clean mechanical inspection usually point to the touch panel itself, so order the replacement early instead of chasing software causes.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.