GE LOGIQ P5: video card cooling fan replacement
What happened
The cooling fan on the graphics card had to be replaced. No error codes were reported in the case.
What was checked
- Find the replacement by the marking on the original fan / video card assembly, not by the ultrasound model.
- Before ordering, confirm frame size, supply voltage, current draw and connector type — in this case a 45 mm, 5 V, 0.40 A fan with a 3-pin connector (figures as reported in the case).
Cause
Most probably the original fan had simply worn out, leaving the graphics card without active cooling; the exact failure mode was not described in the case.
What was done
The engineers matched a replacement to the marking on the fan of the graphics card assembly (DFB500905H T055010SH, 45 mm, 5 V, 0.40 A, 3-pin — as reported in the case), sourced it through an online marketplace and installed it; the impeller had to be trimmed slightly to seat correctly. The problem was solved.
Safety
Disconnect the system from the mains and let it power down before opening, take ESD precautions around the boards, and before returning the unit to use confirm the new fan spins freely with correct airflow direction and pass an electrical safety test.
Lesson: Source cooling fans by the marking on the part itself and check size, voltage, current and connector before ordering — needing to trim an impeller means that check came too late.