GE Logiq P6: trackball completely unresponsive
What happened
On a GE Logiq P6 ultrasound the trackball stopped working outright — rotating it produced no reaction on screen at all.
What was checked
- Roll the trackball rollers with a finger and watch for any cursor reaction on screen, to confirm whether the assembly registers movement at all.
- Open the trackball assembly and inspect the optical sensor pair — in these units the light path can be blocked by dirt.
- With the assembly open, check that the sensor board and its cable are clean and seated properly.
Cause
Not confirmed — the case was still at the diagnostic stage. The engineers' working assumption was dirt or debris blocking the optical sensor, since in their experience there is little inside this assembly that fails outright.
The cause is probable and was not fully confirmed.
What was done
Clean first, replace only if that fails. The suggested fix was to disassemble the trackball, clean the rollers and the optical sensor, and check it works before considering a new part. If replacement does turn out to be necessary, a trackball from a Logiq P5 (the roller type) was reported in the case to fit and to be inexpensive. Since the case did not reach a confirmed repair, the outcome is probable.
Safety
Unplug the system from the mains before opening the operator console, and run the manufacturer's electrical safety tests before returning the scanner to clinical use.
Lesson: Before ordering a replacement for a dead trackball, open and clean it — in these assemblies dirt on the optical sensor is a far more likely culprit than failed electronics.