GE · Logiq P6

GE Logiq P6: trackball completely unresponsive

What happened

On a GE Logiq P6 ultrasound the trackball stopped working outright — rotating it produced no reaction on screen at all.

What was checked

Cause

Not confirmed — the case was still at the diagnostic stage. The engineers' working assumption was dirt or debris blocking the optical sensor, since in their experience there is little inside this assembly that fails outright.

The cause is probable and was not fully confirmed.

What was done

Clean first, replace only if that fails. The suggested fix was to disassemble the trackball, clean the rollers and the optical sensor, and check it works before considering a new part. If replacement does turn out to be necessary, a trackball from a Logiq P5 (the roller type) was reported in the case to fit and to be inexpensive. Since the case did not reach a confirmed repair, the outcome is probable.

Safety

Unplug the system from the mains before opening the operator console, and run the manufacturer's electrical safety tests before returning the scanner to clinical use.

Lesson: Before ordering a replacement for a dead trackball, open and clean it — in these assemblies dirt on the optical sensor is a far more likely culprit than failed electronics.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.