GE Logiq P7: trackball inverted on the vertical axis

What happened

On the reported unit the trackball behaved inverted on the vertical axis: the cursor tracked left and right normally, but rolling the ball up moved the cursor down and vice versa. No error messages were reported.

What was checked

Cause

The engineers found a displaced optocoupler in the trackball mechanism: it sat misaligned relative to the vertical-axis encoder disc, so the rotation was read backwards. In this case it was a hardware fault, not a setting.

What was done

The trackball assembly was disassembled, the optocoupler was re-aligned and re-soldered, and after reassembly the trackball moved correctly on both axes. No parts were replaced.

Safety

After opening the console and soldering inside, run the manufacturer's electrical safety test (protective earth continuity and leakage currents) before returning the scanner to clinical use.

Lesson: An inverted axis on a trackball usually points to a shifted or faulty photo sensor in the mechanism, so check the hardware alignment before spending time on software settings.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.