GE Logiq P7: trackball inverted on the vertical axis
What happened
On the reported unit the trackball behaved inverted on the vertical axis: the cursor tracked left and right normally, but rolling the ball up moved the cursor down and vice versa. No error messages were reported.
What was checked
- The engineers confirmed the behaviour axis by axis: horizontal tracking normal, vertical reversed.
- They searched the system settings for any trackball axis configuration and found nothing relevant, so a settings problem was ruled out.
Cause
The engineers found a displaced optocoupler in the trackball mechanism: it sat misaligned relative to the vertical-axis encoder disc, so the rotation was read backwards. In this case it was a hardware fault, not a setting.
What was done
The trackball assembly was disassembled, the optocoupler was re-aligned and re-soldered, and after reassembly the trackball moved correctly on both axes. No parts were replaced.
Safety
After opening the console and soldering inside, run the manufacturer's electrical safety test (protective earth continuity and leakage currents) before returning the scanner to clinical use.
Lesson: An inverted axis on a trackball usually points to a shifted or faulty photo sensor in the mechanism, so check the hardware alignment before spending time on software settings.