GE Logiq S8: error with any probe connected

What happened

As reported in the case, the system boots normally, but an error message appears as soon as any probe is plugged in, regardless of type.

What was checked

Cause

Not confirmed. The engineers suspected a fault on the BF192 (TX pulser) board, as it is the shared element in the probe path for all connectors; diagnosis was still in progress, so this cause remains probable only.

The cause is probable and was not fully confirmed.

What was done

At the time of the note the repair was still in diagnostics. The intended path was to verify the BF192 with the built-in tests and a visual inspection; any board repair or replacement would follow a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Isolate the system from mains and allow time for internal capacitors to discharge before removing the board, and complete an electrical safety test before returning the unit to patient use.

Lesson: When the same fault appears with every probe, the likely culprit is the shared probe-side electronics rather than the probes themselves, so check the common board early in the diagnosis.