GE · Logiq S8

GE Logiq S8: rainbow color flooding in color Doppler

What happened

On a new GE Logiq S8 with a 3S probe, color Doppler mode floods the image with rainbow-colored artifacts. No error codes are reported in the case.

What was checked

Cause

Probable: a defective 3S probe (elements, lens or cable) generating false Doppler signals. The artifact persisting across power sources and rooms points at the probe, but a single case cannot prove it — the swap has to confirm.

The cause is probable and was not fully confirmed.

What was done

No field repair was possible; a new 3S probe was ordered to replace the suspect one. The diagnosis remains probable until the swap confirms it.

Safety

Disconnect and connect the probe only with the system powered down, and complete the manufacturer's electrical safety tests, including patient-applied-part leakage current, before the swapped probe returns to clinical use.

Lesson: When an imaging artifact follows the machine across power sources and rooms, rule out the probe with a swap before chasing the site's electrical supply.