GE Logiq S8: rainbow color flooding in color Doppler
What happened
On a new GE Logiq S8 with a 3S probe, color Doppler mode floods the image with rainbow-colored artifacts. No error codes are reported in the case.
What was checked
- The engineers reproduced the flooding with the system powered from a UPS and directly from mains
- The engineers reproduced the artifact in a different room, ruling out the local power environment
- The probe was suspected, and a replacement 3S was ordered for a diagnostic swap
Cause
Probable: a defective 3S probe (elements, lens or cable) generating false Doppler signals. The artifact persisting across power sources and rooms points at the probe, but a single case cannot prove it — the swap has to confirm.
The cause is probable and was not fully confirmed.
What was done
No field repair was possible; a new 3S probe was ordered to replace the suspect one. The diagnosis remains probable until the swap confirms it.
Safety
Disconnect and connect the probe only with the system powered down, and complete the manufacturer's electrical safety tests, including patient-applied-part leakage current, before the swapped probe returns to clinical use.
Lesson: When an imaging artifact follows the machine across power sources and rooms, rule out the probe with a swap before chasing the site's electrical supply.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.